| Model Number |
PCB00 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable as the lens was not implanted.If explanted, give date: not applicable as the lens was not implanted and therefore not explanted.(b)(6).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the cartridge tip of the preloaded delivery system, model pcb00, was observed split when the intraocular lens (iol) was advanced through the cartridge.Patient contact was reported.Reportedly, the surgeon removed the delivery system from the patient's eye with the iol still inside the cartridge.The surgery was completed with the back-up lens and there was no patient injury or any other issue.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes, returned to manufacturer on: 01/08/2019.Device returned to manufacturer? yes.Device evaluation: the preloaded delivery system was returned at the manufacturing site for evaluation.Visual inspection showed that the intraocular lens (iol) was stuck at the cartridge tube and tip.The push rod was overriding the lens and deforming the tip.The customer's reported complaint was verified but a product deficiency could not be determine due to the condition of the returned sample.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|