• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX35008UX
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An attempt was made to use one resolute onyx rx coronary drug eluting stent to treat lesion.It was reported that the stent failed to cross the lesion and stent deformation occurred.It is indicated that stent deformation occurred due to calcification of the tortuous vessel.The procedure was completed using a non-mdt stent.No patient injury is reported.
 
Additional Manufacturer Narrative
An attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a heavily calcified non tortuous mid lad lesion exhibiting 90% stenosis.It is indicated that stent deformation occurred due to stent getting caught on the calcification of the vessel.The account reported that the event was related to a sizing issue.The procedure was completed using a smaller non-medtronic stent after further pre-dilation.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8073098
Report Number9612164-2018-03215
Device Sequence Number1721663
Product Code NIQ
UDI-Device Identifier00643169557055
UDI-Public00643169557055
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/29/2019
Device Catalogue NumberRONYX35008UX
Device Lot Number0008556999
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/13/2018
Supplement Date Received by Manufacturer12/18/2018
Initial Report FDA Received Date11/15/2018
Supplement Report FDA Received Date01/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No