MAUDE Adverse Event Report: COOK IRELAND LTD ZILVER 635 BILIARY SELF EXPANDING METAL STENT; FGE CATHETER, BILIARY, DIAGNOSTIC - BILIARY STENT - METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ZILVER 635 BILIARY SELF EXPANDING METAL STENT; FGE CATHETER, BILIARY, DIAGNOSTIC - BILIARY STENT - METAL Back to Search Results
Model Number G50622
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
As reported to customer relations: "when they were going to deploy the stent, they had a lot of resistance and the outer sheath by the handle became separated from the handle.About 2-3 cm from the tip of the catheter, there is a kink on the catheter.The stent is partially deployed (about 1 cm of it).Device was removed, another device was used to complete the procedure successfully.".
 
Additional Manufacturer Narrative
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(6).Cook medical incorporated (cmi) 1025 acuff road, p.O box 4195, bloomington, indiana 47402-4195.Importer site establishment registration number: (b)(4).Device evaluation: the zilbs-635-6-8 device of lot number c1333985 involved in this complaint was returned for evaluation, with the original packaging.With the information provided, a physical examination and document based investigation was conducted.Lab evaluation: the device involved in the complaint was evaluated in the lab on 08 nov 2018.Complaint is confirmed as the failure was verified in the laboratory.The flexor was separated.It may be noted that a project (b)(4) has been completed and addresses the issue of outer sheath separation from the handle.The lot involved in this complaint was manufactured prior to implementation of this project.Document review and ifu review: prior to distribution all zilbs-635-6-8 devices are subject to visual inspection and functional checks to ensure device integrity.A review of the relevant manufacturing records (b)(4) revealed no discrepancies that could have contributed to this complaint.Upon review of complaints, this failure mode has not occurred previously with this lot number (b)(4).Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number (b)(4).Ifu review: there is no evidence to suggest that the customer did not follow the instructions for use.Root cause review: a definitive root cause could not be determined as the circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to the resistance encountered when advancing the stent.Summary: complaint is confirmed as the failure was verified in the laboratory.The flexor was separated.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
As reported to customer relations: "when they were going to deploy the stent, they had a lot of resistance and the outer sheath by the handle became separated from the handle.About 2-3 cm from the tip of the catheter, there is a kink on the catheter.The stent is partially deployed (about 1 cm of it).Device was removed, another device was used to complete the procedure successfully.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZILVER 635 BILIARY SELF EXPANDING METAL STENT
Common Device Name
FGE CATHETER, BILIARY, DIAGNOSTIC - BILIARY STENT - METAL
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key8073138
Report Number3001845648-2018-00534
Device Sequence Number14044300
Product Code FGE
UDI-Device Identifier10827002506227
UDI-Public(01)10827002506227(17)200310(10)C1333985
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K020788
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2020
Device Model NumberG50622
Device Catalogue NumberZILBS-635-6-8
Device Lot NumberC1333985
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/18/2018
Event Location Hospital
Initial Date Received by Manufacturer 10/18/2018
Supplement Date Received by Manufacturer10/18/2018
Initial Report FDA Received Date11/15/2018
Supplement Report FDA Received Date12/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-