• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH FLEXIVA (TM) HIGH POWER SINGLE-USE LASER FIBER; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH FLEXIVA (TM) HIGH POWER SINGLE-USE LASER FIBER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M0068403920
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Investigation results: one flexiva 365 laser fiber was received for evaluation.Visual examination revealed that the exposed glass fiber tip measured approximately 3.5 mm and appeared unused.Additional examination under magnification confirmed that the tip was unused.There was no damage along the body of the fiber.Visual examination revealed that light cannot travel to the fiber face within the sma connector to illuminate it, this indicated that the fiber was broken within the connector.The condition of the returned device would cause the device to emit insufficient energy transmission during ablation thereby confirming the event.The noted damages indicate difficulty was experienced during the procedure and was likely due to anatomical or procedural factors such as tortuous anatomy or maneuvering of the device.Therefore, a review and analysis of all available information indicated that the most probable root cause is operational context.A review of the device history record (dhr) was performed; no anomalies were noted.A search of the complaint database revealed that no similar complaints exist for the specified lot.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that a flexiva 365 laser fiber was used during a ureteroscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure and outside the patient, the laser fiber did not work upon pedal activation and it was noted to be burned.The procedure was completed with another flexiva 365 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.Note: this event has been deemed an mdr-reportable event based on investigation results which revealed that the laser fiber was broken within the connector.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIVA (TM) HIGH POWER SINGLE-USE LASER FIBER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
MDR Report Key8073164
Report Number3005099803-2018-02626
Device Sequence Number2165605
Product Code GEX
UDI-Device Identifier08714729808183
UDI-Public08714729808183
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K100078
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/18/2020
Device Model NumberM0068403920
Device Catalogue Number840-392
Device Lot Number0000005090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2018
Initial Report FDA Received Date11/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age46 YR
Patient Weight74
-
-