| Model Number |
MMT-722NAS |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.Customer's blood glucose value was 151 mg/dl at the time of the incident.Customer states that the drive support cap was protruded.The customer was advised the insulin pump has to be replaced and revert to back-up plan per health care provider's instruction.Customer will return the device for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime test, excessive no delivery test and occlusion test or verify compromised force sensor system alarm due to motor error alarm.
|
| |
|