PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4002C1245
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was undamaged.The embolization coil was intact with the pusher assembly.The embolization coil had offsets coil winds along length of the coil.Conclusions: evaluation of the returned pc400 revealed a device with offset coil winds.If the pc400 introducer sheath is not properly aligned with the parent device when attempting to advance, resistance may be experienced.If the device is forcefully advanced while experiencing resistance, damage such as offset coil winds may occur.These offset coil winds likely contributed to the resistance experienced during functional testing.The non-penumbra microcatheter was not returned for evaluation.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra coil 400s (pc400s).During the procedure, the physician successfully placed four pc400s using a non-penumbra microcatheter.The physician then felt resistance and was unable to advance another pc400 through the hub of the microcatheter and therefore the pc400 was removed.The procedure was completed using a new pc400, additional coils, and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8073275
Report Number3005168196-2018-02262
Device Sequence Number1721728
Product Code HCG
UDI-Device Identifier00814548010649
UDI-Public00814548010649
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/18/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/24/2024
Device Catalogue Number4002C1245
Device Lot NumberF71318
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2018
Initial Report FDA Received Date11/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-