• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUETRACK
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Codes Nausea (1970); Dizziness (2194)
Date of Event 10/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) internal report # (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-68- user did not insert code key.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for code eci meter.During the call on (b)(6) 2018, a check strip test was performed.Customer removed and properly reinserted the code chip and test strip and the error message still displayed.A new vial of test strips was used, same lot number, and the error message still displayed.The test strip manufacturer's expiration date is 12/21/2020.At the time of the call, the customer was feeling ill and reported symptoms of dizziness and nausea.Customer stated he would go to the hospital.No medical attention was reported at the time of the call.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8073426
Report Number1000113657-2018-01161
Device Sequence Number1616332
Product Code NBW
UDI-Device Identifier00021292001797
UDI-Public(01)00021292001797
Combination Product (Y/N)N
Initial Reporter StateTN
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 11/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/21/2020
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRV5276
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2018
Initial Report FDA Received Date11/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-