| Catalog Number |
60P040 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/09/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Report source:(b)(6).
|
| |
|
Event or Problem Description
|
|
Information was received indicating that nine days following cannula change out to a smiths medical portex® bivona® uncuffed tracheostomy tube a breakage to the flange was noted.Subsequently, a tube change out was required with no reported adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
One tracheostomy was returned for evaluation.Visual inspection found the device to have two splits on the flange eyelet.Device underwent testing by making a small cut in a neckflange eyelet with a blade into two pieces.As a result, the customer complaint was confirmed.The most probable root cause is that damaged occurred after the product left the shm facility.
|
| |
|