MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. BONE SCREW SELF-TAPPING 6.5 MM DIA. 30 MM LENGTH; PROSTHESIS HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. BONE SCREW SELF-TAPPING 6.5 MM DIA. 30 MM LENGTH; PROSTHESIS HIP Back to Search Results
Model Number N/A
Device Problem Material Twisted/Bent (2981)
Patient Problem No Information (3190)
Event Date 10/10/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source- (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that when the product was trying to be placed, screw was bent from the neck part and became useless.Attempts have been made and additional information on the reported event is unavailable at this time.No further information is available.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D4 - udi # - (b)(4).Reported event was confirmed by review of a trilogy bone screw (lot 63980990) which was returned for evaluation.As returned, the device is bent and exhibits thread damage.Surface damage is seen on the spherical radius and the shank.This damage appears to be from impingement.The major and minor thread diameters are conforming to print specifications.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW SELF-TAPPING 6.5 MM DIA. 30 MM LENGTH
Type of Device
PROSTHESIS HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
MDR Report Key8074128
MDR Text Key127368606
Report Number0002648920-2018-00826
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
PMA/PMN Number
K934765
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00625006530
Device Lot Number63980990
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2019
Was the Report Sent to FDA? No
Date Manufacturer Received07/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A