Model Number N/A |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Information (3190)
|
Event Date 10/10/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source- (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that when the product was trying to be placed, screw was bent from the neck part and became useless.Attempts have been made and additional information on the reported event is unavailable at this time.No further information is available.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.D4 - udi # - (b)(4).Reported event was confirmed by review of a trilogy bone screw (lot 63980990) which was returned for evaluation.As returned, the device is bent and exhibits thread damage.Surface damage is seen on the spherical radius and the shank.This damage appears to be from impingement.The major and minor thread diameters are conforming to print specifications.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|