MEDTRONIC SINGAPORE OPERATIONS CAPSURE SENSE; ELECTRODE, PACEMAKER, PERMANENT Back to Search Results
Model Number 407458
Medical Device Problem Codes Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
Health Effect - Clinical Code Bradycardia (1751)
Date of Event 09/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported by the patient that the patient experienced bradycardia three days post implant and it was discovered that a lead was not "attached".Upon follow-up with the patient's following physician, it was indicated that the right ventricular (rv) lead had high thresholds.It was noted that a chest x-ray revealed the rv lead had a loss of slack and a microdislodgement was suspected.The rv lead was planned to be revised and remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product event summary: the full lead was returned and analyzed no anomalies were found.Correction: if information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was further reported that the right ventricular (rv) lead was explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE SENSE
Common Device Name
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SG  486056
MDR Report Key8074191
Report Number3008973940-2018-02287
Device Sequence Number2262332
Product Code DTB
UDI-Device Identifier00643169448957
UDI-Public00643169448957
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P830061
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source consumer,health professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/08/2019
Device Model Number407458
Device Catalogue Number407458
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2018
Initial Date Received by Manufacturer 10/22/2018
Supplement Date Received by Manufacturer11/15/2018
12/05/2018
Initial Report FDA Received Date11/15/2018
Supplement Report FDA Received Date11/20/2018
01/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
A2DR01 IPG, 5076-52 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
-
-