BOSTON SCIENTIFIC NEUROMODULATION; PROBE, RADIOFREQUENCY LESION Back to Search Results
Model Number CSK-TC10
Medical Device Problem Code Insufficient Heating (1287)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/01/2018
Type of Reportable Event Malfunction
Event or Problem Description
Device analysis performed on the returned electrodes found that the shaft of the electrode was separated from the hub.The damage was likely due to unintended excessive external mechanical force.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
PROBE, RADIOFREQUENCY LESION
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
MDR Report Key8074563
Report Number3006630150-2018-61840
Device Sequence Number1733025
Product Code GXI
UDI-Device Identifier00813250010107
UDI-Public00813250010107
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K050084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCSK-TC10
Device Catalogue NumberCSK-TC10
Device Lot NumberP691
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2018
Initial Report FDA Received Date11/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-