| Model Number |
2067L |
| Medical Device Problem Code |
Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: it was found that, when the rf head was connected to a known working programmer, it would cause the programmer to power off.The rf head was forwarded for repair.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The radio frequency (rf) programmer head was returned for inspection, and subsequently tested out of specification.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis:additional analysis confirmed that when the radiofrequency (rf) head was plugged into a programmer, the programmer would power down.The rf head failed 30 out of 44 functional tests.The rf head cable was replaced.The rf upper case was replaced due to a cracked lens.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
| |
|