ARTHREX, INC. DISSECTOR, CURVED 4.0MM X 13CM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
| Model Number |
DISSECTOR, CURVED 4.0MM X 13CM |
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Visual evaluation revealed metal burrs and scuff markings along the od of the serrations and outer tube window of both ar-8400cds dissectors, consistent with sites of debris production.Returned, new devices of the same part/lot number combination did not display the aforementioned characteristics.This event is most likely caused by excessive bending forces applied to the device during use.
|
| |
|
Event or Problem Description
|
|
It was reported by the sales rep that during a meniscal repair upon insertion the ar-8400cds was producing metal shaving.The surgeon changed the ar-8400cds same lot number (10212946) and the problem occurred again.The surgeon used a shaver to suction the metal shaving, the sales rep was unsure if all the metal shaving were removed.No x-ray was taking after the procedure was complete.The case was completed using a third ar-8400cds with no additional issue.
|
| |
|