| Model Number |
4CFS |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, there was a separation of the cuff from the rest of the cannula.The customer reported that there was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one sample of device was received.It was observed that the flange was detached from the cannula, along with the outer cannula.A visual inspection was performed and it was observed the outer cannula does not present a cuff which is according to manufacturing since this is a cuffless tracheostomy tube.A reported event was observed.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
| |
|