MAUDE Adverse Event Report: EPIC INTERNATIONAL (THAILAND) CO., LTD. SMARTEZ PUMP; ELASTOMERIC INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EPIC INTERNATIONAL (THAILAND) CO., LTD. SMARTEZ PUMP; ELASTOMERIC INFUSION PUMP Back to Search Results
Model Number 481062
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigations on similar defective samples that were returned, the no flow events were due to kinked administration tube arising from clamping of the device.
 
Event or Problem Description
Two (2) smartez pumps (250 ml volume; 175 muhr) filled with vancomycin 1 g/0.9% sodium chloride would not infuse.Smartez pump lot: s8h11.Therapy start date: (b)(6) 2018.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTEZ PUMP
Common Device Name
ELASTOMERIC INFUSION PUMP
Manufacturer (Section D)
EPIC INTERNATIONAL (THAILAND) CO., LTD.
hemaraj eastern seaboard indus
500/73 moo 2, tasit
pluakdaeng, rayong 21140
TH  21140
MDR Report Key8074954
Report Number3012429465-2018-00001
Device Sequence Number195067
Product Code MEB
UDI-Device Identifier28859299101068
UDI-Public(01)28859299101068
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K151650
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/28/2021
Device Model Number481062
Device Lot NumberS8H11
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/06/2018
Initial Report FDA Received Date11/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-