If additional information should become available, a supplemental medwatch will be submitted accordingly.(b)(4).Investigation summary: the complaint device was discarded by the customer therefore the device is not available for a physical evaluation.This complaint is not confirmed.Without physically evaluating the device, a definite root cause cannot be determined.No nonconformances were identified for this part number (210816), lot number (l775627) combination.At this point in time, no corrective action is required, and no further action is warranted.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
|
It was reported by the sales rep via phone that during an acl reconstruction procedure, it was observed that the rigidfix 3.3mm femoral cross pin kit drill bit device that goes through the cannulated portion of the device cold-welded together.The sales rep stated that due to this the guide would not stay in the patient.The case was completed with another like-device with no patient harm or delay.The device was discarded by the customer.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|