Catalog Number 31302170S
Health Effect - Clinical Codes Pain (1994); Ambulation or Postural Difficulties (2544)
Date of Event 10/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available it will be provided on a supplemental report.
 
Event or Problem Description
Cut out of u lag screw occurs.Patient felt pain on his hip for several weeks and difficult to walk.Then, when falling on (b)(6), visited in the hospital and an x - ray film was taken, it turned out that the lag screw had cut out and reached the acetabular.There is no dislocation of the pieces of the bony skull, and only the lag screw has cut out.As a measure after occurrence, on (b)(6) replaced 120 ° gamma nail.
 
Additional Manufacturer Narrative
The reported event that trochanteric nail kit, ti gamma3® ø12x170mm x 130° was alleged of issue (implant revision / removal without implant breakage) could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of manufacturing or design related problems were found during the investigation.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
Cut out of u lag screw occurs.Patient felt pain on his hip for several weeks and difficult to walk.Then, when falling on (b)(6), visited in the hospital and an x - ray film was taken, it turned out that the lag screw had cut out and reached the acetabular.There is no dislocation of the pieces of the bony skull, and only the lag screw has cut out.As a measure after occurrence, on (b)(6) replaced 120 ° gamma nail.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCHANTERIC NAIL KIT, TI GAMMA3® Ø12X170MM X 130°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
MDR Report Key8074995
Report Number0009610622-2018-01382
Device Sequence Number2325421
Product Code HSB
UDI-Device Identifier04546540375148
UDI-Public04546540375148
Combination Product (Y/N)N
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date03/31/2021
Device Catalogue Number31302170S
Device Lot NumberK0D136
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/18/2018
Supplement Date Received by Manufacturer11/29/2018
Initial Report FDA Received Date11/15/2018
Supplement Report FDA Received Date12/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
-
-