| Catalog Number |
3542911 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no device history record (dhr) review could be performed.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) caucasian female patient underwent a primary breast augmentation with mentor siltex contour profile moderate 225/250cc saline breast implants which the right side deflated after implantation.Patient woke up and visibly observed the right breast deflated.As a result, patient is scheduled for removal on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
On 12/20/2018, the mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary upon receipt to mentor, the device contained no fluid.Orange material was observed within the device.White and orange material were observed on the shell surface.Visual evaluation of the device revealed a rent within an area of siltex cracking on the posterior view, measuring 1.0 cm.No other anomalies were found.Mentor products are 100% visually inspected prior to release in addition to thorough in-process testing during several stages of the manufacturing process.Mentor product analysis lab concluded that the rent and siltex cracking occurred sometime subsequent to the removal of the device from its protective packaging.The complaint of deflation was confirmed since a rent was found on the device.However, at this point it is not possible to determine the root cause of such damage or the origin of the orange and white material observed.Mentor products are 100% visually inspected prior to release in addition to thorough in-process testing during several stages of the manufacturing process.Mentor product analysis lab concluded that the rent and siltex cracking occurred sometime subsequent to the removal of the device from its protective packaging.All the implants are 100% inspected before leaving the facility.There is no evidence that the issue is related with manufacturing breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|