U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000
Device Problems Insufficient Cooling (1130); Inadequacy of Device Shape and/or Size (1583); Improper or Incorrect Procedure or Method (2017)
Patient Problems Alteration In Body Temperature (2682); Patient Problem/Medical Problem (2688)
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The device was not returned.
 
Event Description
It was reported that the arctic sun device had low flow and the patient was slow to cool. The complainant noted that the facility ran out of appropriate sized pads for the patient and was using three universal pads. The patient had come into the facility for a drug overdose and was currently on versed, propofol, and fentanyl. The patient's temperature was 37. 3c with a target temperature of 33c. The water temperature was 4. 8c and the flow rate was 1. 6 lpm. The patient was set to reach the target within four hours and had been on the device for three hours at the time of the call. Upon call back approximately 2 hours later, the complainant replaced the three universal pads with a full set of pads borrowed from another facility. The patient's temperature was 37. 1c and the water temperature was 5. 1c. The flow rate had increased to 2. 4 lpm. Ms&s discussed causes of heat generation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key8075107
MDR Text Key127291959
Report Number1018233-2018-05438
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 11/15/2018 Patient Sequence Number: 1
-
-