| Catalog Number |
18955090S |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
At (b)(6) 2018, operation with t2scn was done.After the operation, the end cap was came out and the most distal condyle screw was backed out.Therefore, at (b)(6) 2018, the end cap and condyle screw was removed and new same devices were implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that condyle screw t2 femur ø5x60 mm was alleged of implant migration could not be confirmed.Although x-rays were repeatedly requested but not provided until time of investigation closure and thus, the reported event of unintended movement of condyle screw after implantation could not be verified.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.The device inspection revealed the following: we received a condyle screw t2 femur ø5x60 mm for evaluation.Detailed visual inspection revealed clear signs of usage.The windings show material abrasion.The rim of the hexagon shows clear signs of damage as well.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.
|
| |
|
Event or Problem Description
|
|
At (b)(6) 2018, operation with t2scn was done.After the operation, the end cap was came out and the most distal condyle screw was backed out.Therefore, at (b)(6) 2018, the end cap and condyle screw was removed and new same devices were implanted.
|
| |
|