Manufacturing date: added, device evaluated by mfg: updated, summary attached: updated, expiration date: added.The device history record review confirms that the device met all material, assembly and performance specifications.The microcatheter and torque device used with the guidewire were not returned.Visual inspection revealed that the guidewire was received in two fragments.Magnified examination of the fracture site showed the core wire was fractured.The guidewire was bent (kinked) at the fracture site and the guidewire was also bent just proximal and distal to the fracture site as well.The device was kinked at approximately 7.5cm from its distal end and bent along its polymer sleeve length.The ptfe (polytetrafluoroethylene) coating was also found to be scraped close to the fracture site of the guidewire.Functional testing could not be performed due to the damaged condition of the returned guidewire.The bending and fracture on the guidewire appeared to be due to excessive manipulation.Information available indicated that resistance was felt when pulling the wire out from the dispenser hoop and the wire separated following removal.Analysis of the device revealed bending at the fracture site which is an indication of excessive manipulation.From the condition of the fracture, it appears that the guidewire was kinked, scraped, and then separated due to excessive force applied to overcome the reported resistance during removal from the dispenser hoop.Therefore based on all of the information available, an assignable cause of handling damage was assigned to the observed and reported damages.
|