Model Number 10014916
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: mz1000-07, therapy date: (b)(6) 2018.The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
 
Event or Problem Description
The customer reported a tubing leak at an unspecified location.It was reported that the male adapter was changed and the set was tightened.There was no report of patient harm.
 
Event or Problem Description
The customer reported a tubing leak at an unspecified location.It was reported that the male adapter was changed and the set was tightened.There was no report of patient harm.
 
Additional Manufacturer Narrative
The customer¿s report of a tubing leak was confirmed.Visual inspection of the set noted a crack along the female luer component of the set, which was found to be cracked along the top and continuing along the length of its luer.The crack was observed to be opposite the injection gate on the luer component.No tool markings were observed around the damaged area.Functional testing confirmed leaking from the damaged female luer.The cause of the leak was due to an observed crack along the set's female luer.The root cause of the observed damage was as a combination of material degradation during the luer component's supplier process and manufacturing factors (solvent and over-torque).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SYRINGE MODULE SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key8078283
Report Number9616066-2018-02252
Device Sequence Number13893105
Product Code FPA
UDI-Device Identifier10885403233920
UDI-Public10885403233920
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K811885
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10014916
Device Catalogue Number10014916
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/16/2018
Supplement Report FDA Received Date02/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
8110, 8015, PRI TUBING, SYRINGE, TD (B)(6) 2018
-
-