| Model Number |
9735602 |
| Medical Device Problem Code |
Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.Missing device manufacture date.No parts have been returned to the manufacturer for analysis.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of procedure.It was reported that the exams would not load after selecting the correct unstructured option.After rebooting several times the exams loaded.There was no patient present.
|
| |
|
Additional Manufacturer Narrative
|
|
A medtronic representative went to the site to test the equipment.Testing revealed that the system was functioning as intended.The system passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
|
| |
|