| Catalog Number |
20392 |
| Medical Device Problem Codes |
Expiration Date Error (2528); Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Asp complaint ref #: (b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported using expired cidex® opa test strips to test their cidex® opa solution and the instruments that were cleaned in the tested cidex® opa solution were released and used on patients.They stated they used the test strips because the instructions for use (ifu) states the strips can be used for 90-days after opening the bottle, and the cidex® opa strips came out positive (or purple) indicating the cidex® opa solution meets the minimum effective concentration (mec).There was no report of infection, injury or harm to patient(s) associated with this issue.The customer was advised to discard any remaining cidex® opa test strips and to follow the instructions for use (ifu) which states the unused cidex® opa test strips should be discarded after the expiration date on the outside of the bottle even if it has not reached the 90-day open date.As a matter of policy, advanced sterilization products (asp) has decided to report cases where a customer uses expired product and releases it for use on patients since they cannot guarantee high-level disinfection has been achieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Asp investigation summary: the investigation included a review of the batch history record, complaint trending by lot number, and system risk analysis (sra).The batch history record was reviewed and test specifications for product release were met.No issues were observed that would contribute to the complaint.Complaint trending by lot number was reviewed for the prior six months from open date and no significant trend was observed.The sra indicates the risk associated with exposure to biohazardous, pathogenic or infectious material is "low." functional analysis was not performed.The customer disposed of the expired test strips and stated they will not use any test strips beyond the expiration date on the outside of the bottle.The likely assignable cause of this issue is due to user error and failure to follow the instructions for use.The issue will continue to be tracked and trended.Asp complaint ref #: (b)(4).
|
| |
|
Search Alerts/Recalls
|