| Catalog Number |
352424 |
| Medical Device Problem Code |
Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The sample has not yet been received and the investigation is ongoing at this time.A follow-up report will be submitted when the results of the investigation become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: the line detached above the filter which allowed air into the line causing a steady stream of bubbles to come into the drip chamber, inhibiting the line to drip properly.No injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No samples were returned for evaluation in connection with this incident.Multiple unsuccessful attempts were made to obtain a sample.Without the actual device, a thorough investigation could not be performed.Review of the discrepancy management system (dsms) database performed for the reported lot number revealed no abnormalities or non-conformances noted during in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If a sample and/or any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|