| Model Number |
801041 |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The srt observed the display to not turn on when the pump was plugged in.After tapping on the screen, the display operated normally.
|
| |
|
Event or Problem Description
|
|
The service repair technician (srt) reported that during routine testing of the device at the service center, the pump had intermittent display operation.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint was confirmed.The service repair technician (srt) will replace the display assembly.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|