| Catalog Number |
367671 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd vacutainer® sodium heparinn (nh) blood collection tubes the gel was not separating properly.There was no report of exposure, injury, or medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd had reached out to the customer and provided bd's recommendations for specimen collection handling and processing in relation to the issue they were observing.The customer acknowledged this information and has indicated that the issue has been resolved.Investigation conclusion: as no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.Root cause description: the most likely root cause of the issue has been determined to be related to specimen collection handling and processing.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd vacutainer® sodium "heparin" (nh) blood collection tubes the gel was not separating properly.There was no report of exposure, injury, or medical intervention.
|
| |
|