Catalog Number 367671
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd vacutainer® sodium heparinn (nh) blood collection tubes the gel was not separating properly.There was no report of exposure, injury, or medical intervention.
 
Additional Manufacturer Narrative
Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd had reached out to the customer and provided bd's recommendations for specimen collection handling and processing in relation to the issue they were observing.The customer acknowledged this information and has indicated that the issue has been resolved.Investigation conclusion: as no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.Root cause description: the most likely root cause of the issue has been determined to be related to specimen collection handling and processing.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that the bd vacutainer® sodium "heparin" (nh) blood collection tubes the gel was not separating properly.There was no report of exposure, injury, or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SODIUM HEPARINN (NH) BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key8078500
Report Number1917413-2018-03773
Device Sequence Number13886777
Product Code JKA
UDI-Device Identifier50382903676717
UDI-Public50382903676717
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2019
Device Catalogue Number367671
Device Lot Number7340869
Initial Date Received by Manufacturer 10/23/2018
Supplement Date Received by Manufacturer10/23/2018
Initial Report FDA Received Date11/16/2018
Supplement Report FDA Received Date01/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-