| Model Number |
11088483 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 10ml syringe, ref (b)(4), lot 8060885, exp (b)(6) 2021, 0.9% sodium chloride injection.The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the filter cracked and leaked during patient use in nicu room 2.There was no report of patient harm.
|
| |
|
Event or Problem Description
|
|
The customer reported that the filter cracked and leaked during patient use in nicu room 2.There was no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report of a filter crack was not confirmed as the received set did not have a filter and this model is not manufactured with a filter.The customer¿s report of a leak was confirmed.Visual inspection showed that the female luer had a lateral crack.Closer inspection under magnification showed no tool marks or signs of over torque.The cause of the leak is a crack in the female luer component.The root cause of the crack could not be identified.
|
| |
|