• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1550225-15
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported material rupture.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.The other xience sierra device referenced is being filed under a separate medwatch mfr number.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion located in the right coronary artery that was heavily calcified.After rotablation was performed, balloons were used for predilatation.The 3.0 x 33 mm xience sierra stent delivery system (sds) was advanced to the lesion without resistance, however, during the first inflation, the stent expanded but contrast was leaking and it was not possible to inflate the balloon.The xience sierra was removed from the anatomy and an attempt was made to inflate the balloon.The balloon expanded; however, there were multiple holes in the balloon.Then, the 2.25 x 15 mm xience sierra was advanced to the lesion but ruptured below rated burst pressure during the first inflation.Both sds balloons ruptured prematurely post-rotablation.A 3 x 32 mm non-abbott stent was deployed to complete the procedure.The devices were prepared according to the instructions for use and air aspiration was performed outside the anatomy prior to use without any issue.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8079906
Report Number2024168-2018-08893
Device Sequence Number215638
Product Code NIQ
UDI-Device Identifier08717648227011
UDI-Public08717648227011
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/22/2019
Device Catalogue Number1550225-15
Device Lot Number8051441
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2018
Initial Report FDA Received Date11/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Patient Weight97
-
-