• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DYONICS 4.0MM ELITE STONECUTTER BURR; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DYONICS 4.0MM ELITE STONECUTTER BURR; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200728
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/23/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was overheating.No patient injury was reported.
 
Additional Manufacturer Narrative
One (b)(4) dyonics 4.0mm elite stonecutter burr device returned.The product sluff chamber and adapter body has symptoms of melting.Noisy/grinding and overheating/melted material conditions have been found to be consistent with the device being tested/run without or with inadequate suction/irrigation.This device is not intended to be engaged without suction either.Irreversible damage may occur to blades or burrs if they are run without the flow of irrigation (dry).This can cause bone or tissue fragments to not be excised efficiently.Periodic irrigation of the blade is recommended to provide adequate cooling of the blade and to prevent accumulation of excised materials in the surgical site.Ensure that suction of 128 mmhg minimum is flowing while the instrument is running.There is physical evidence of bone/tissue and/or metal shavings not excising efficiently.As a result, binding and melting occurred.Root cause related to the manufacture of the device was not confirmed.Correction.
 
Event or Problem Description
It was reported that the dyonics was overheating.A backup device was available to complete the procedure with no delay and no patient injuries.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYONICS 4.0MM ELITE STONECUTTER BURR
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8080917
Report Number1219602-2018-01577
Device Sequence Number1579329
Product Code HAB
UDI-Device Identifier03596010578068
UDI-Public03596010578068
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72200728
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2018
Initial Date Received by Manufacturer 10/30/2018
Supplement Date Received by Manufacturer01/10/2019
Initial Report FDA Received Date11/16/2018
Supplement Report FDA Received Date01/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-