Visual inspection and functional testing could not be performed because the device in question has not been returned for evaluation.Thus, the complaint could not be verified and a root cause could not be determined with confidence.It is difficult to perform an accurate evaluation of a failure without having the failed product.As such the complaint is being closed without conclusion.However, if the product is returned in the future the complaint can be reopened and evaluated.A review of relevant clinical/medical information in the reported issue, inclusive of technique and patient information, to include, but not limited to: patient information surgical procedure/post-operative care review device labeling (including technique guides, ifus, etc.) conclusion: it is unknown which part of the device was retained.If the retained suture-fix ul anr material is par of the anchor then it is comprised of a biocompatible material but if it¿s a piece of the instrument then its only been approved for limited tissue/bone exposure.Based on the product evaluation the root cause of the suturefix ul anr breakage is contributed to a user vs procedural event.Since, the location of the retained suturefix ul anr is unknown, we are unable to determine if corrosion/ or micro-motion is possible.The future impact to the patient cannot be determined.No further clinical/medical assessment is warranted at this time.Approved by (b)(6) md on (b)(4) 2018.Conclusion: it is unknown which part of the device was retained.If the retained suture-fix ul anr material is par of the anchor then it is comprised of a biocompatible material but if it¿s a piece of the instrument then its only been approved for limited tissue/bone exposure.Based on the product evaluation the root cause of the suturefix ul anr breakage is contributed to a user vs procedural event.Since, the location of the retained suturefix ul anr is unknown, we are unable to determine if corrosion/ or micro-motion is possible.The future impact to the patient cannot be determined.No further clinical/medical assessment is warranted at this time.Approved by (b)(6) md on (b)(4) 2018.
|