| Model Number |
METRIX AIR |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.(b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer states they will call back and do not have the meter with them at this time.Provided customer call back number and reference number.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results.The customer is concerned with all of the results obtained.The expected fasting blood glucose test result range is 110 to 128mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the bedroom.During the call on (b)(6) 2018, a back to back blood test was performed non-fasting by the customer and produced test results of 214 and 181mg/dl using true metrix air meter.The test strip lot manufacturer's expiration date is 10/20/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly) result 1:217mg/dl date: (b)(6) 2018 time:1:36pm fasting; result 2:238mg/dl date: (b)(6) 2018 time:1:35pm fasting; result 3:170mg/dl date: (b)(6) 2018 time:1:34pm fasting; result 4:221mg/dl date:(b)(6) 2018 time:11:16am fasting; result 5:256mg/dl date: (b)(6) 2018 time:12:49pm fasting.
|
| |
|
Search Alerts/Recalls
|