MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number SMXP130RH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
 
Event or Problem Description
It was reported that (b)(6)-year-old female patient underwent a primary reconstruction surgery with a artoura breast tissue expander 475cc device that deflated after implantation.Deflation of the left tissue expander device was reported.As a result, the patient underwent explantation on (b)(6) 2018.
 
Additional Manufacturer Narrative
On (b)(6) 2018, mentor completed the investigation on the suspect medical device.Upon receipt by mentor, the device contained blue fluid.The device appeared intact upon initial examination.Leak testing, per mentor procedures, revealed no leakage sites in the device which may have contributed to the reported complaint.No anomalies were discovered.Complaint was not confirmed.Mentor products are 100% visually inspected prior to release in addition to thorough in-process testing during several stages of the manufacturing process.A root cause failure analysis will not be conducted since the investigation of the reported failure could not confirm that an actual failure of the device to conform to expected performance had occurred.Because the risk of device deflation cannot be eliminated, mentor addresses this risk in the product insert data sheet (pids) by stating that these devices should not be considered lifetime devices.(b)(4).
 
Additional Manufacturer Narrative
On 11/17/2018, the mentor product analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key8081413
Report Number1645337-2018-06892
Device Sequence Number19422393
Product Code LCJ
UDI-Device Identifier00081317024316
UDI-Public00081317024316
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/09/2022
Device Catalogue NumberSMXP130RH
Device Lot Number7556304
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2018
Initial Date Received by Manufacturer 11/01/2018
Supplement Date Received by Manufacturer11/17/2018
12/11/2018
Initial Report FDA Received Date11/16/2018
Supplement Report FDA Received Date12/03/2018
12/12/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-