• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUTONIX, INC LUTONIX 035 DRUG COATED BALLOON PTA CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUTONIX, INC LUTONIX 035 DRUG COATED BALLOON PTA CATHETER Back to Search Results
Model Number 9004
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Reocclusion (1985)
Date of Event 02/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis: the sample was not returned from the user facility; therefore, a device evaluation is unable to be performed.A lot history review revealed 4 additional reocclusion complaints were associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual sample was not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was not related to the study device or procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.
 
Event or Problem Description
It was reported through a clinical registry that during the index procedure, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter was used to treat the target lesion located in the left distal superficial femoral artery (sfa).Approximately 18 months after the index procedure, the patient¿s target lesion was reportedly reoccluded via duplex ultrasound (dus) imaging.A clinically driven revascularization was performed and the health care professional (hcp) deemed it was successful.The investigator assessed that the event was not related to the study device or procedure.The sample was discarded by the user facility and is not available for evaluation.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
H10: manufacturing review: lot history review revealed there are a total of (b)(4) complaints reported for lot number gfap2162 and the total of five complaints were reported for the allegation of reocclusion.The dhr review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned by the facility for further evaluation.It is known that approximately (b)(4) months after the index procedure, the patient¿s target lesion was reportedly reoccluded via duplex ultrasound (dus) imaging.A clinically driven revascularization was performed and the health care professional (hcp) deemed it was successful.The investigator assessed that the event was not related to the study device or procedure.Approximately (b)(4) months after the index procedure, the patient¿s target lesion was reportedly reoccluded via duplex ultrasound (dus) imaging.Label review: based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.H10: g4 h11: b5, b6, d3, d5 f12, g1, g2, h6 h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through a clinical registry that during the index procedure, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter was used to treat the target lesion located in the left distal superficial femoral artery (sfa).Approximately 18 months after the index procedure, the patient¿s target lesion was reportedly reoccluded via duplex ultrasound (dus) imaging.A clinically driven revascularization was performed and the health care professional (hcp) deemed it was successful.The investigator assessed that the event was not related to the study device or procedure.The sample was discarded by the user facility and is not available for evaluation.No adverse patient effects were reported.New information: it was reported through a clinical registry that approximately 33 months post index procedure, reocclusion was identified via imaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED BALLOON PTA CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
LUTONIX, INC
9409 science center dr
new hope MN 55428
MDR Report Key8081966
Report Number3006513822-2018-00268
Device Sequence Number295664
Product Code ONU
UDI-Device Identifier00801741123306
UDI-Public(01)00801741123306
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,stu
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Expiration Date02/19/2018
Device Model Number9004
Device Catalogue NumberLX3513061505F
Device Lot NumberGFAP2162
Was Device Available for Evaluation? No
Event Location Hospital
Initial Date Received by Manufacturer 10/22/2018
Supplement Date Received by Manufacturer09/20/2019
Initial Report FDA Received Date11/16/2018
Supplement Report FDA Received Date09/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
-
-