(b)(4).Device evaluation by manufacturer: a field service engineer (fse) contacted the customer over the phone to address the reported event.The customer reported that they found the screw and put it back on the top of the probe.They ran calibration and quality controls without any issues.The instrument was operating as expected.Upon follow up, the customer reported that they were running without any issues.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 01-oct-2017 through aware date (b)(4) 2018.There were no similar complaints found during this time.The g8 operator's manual under chapter 5 maintenance procedures under section 5.10 sampling needle replacement: open the left-side door and loosen two needle cover screws indicated in fig.5-11.You do not need to remove the screws.Open the door below the display and loosen one needle screw indicated in fig.5-11.Grasp the needle cover, push it inward while taking care not to break it, and remove it from the needle screw indicated in 3) above.Once you verify that the cover has been removed from the screws shown in fig.5-11, remove the cover by pulling it straightforward.You will see the sampling needle unit back in the middle.Grasp the upper part of the sampling needle unit by hand and slowly pull the unit forward as much as possible.Since a small volume of reagent will spill during replacement, place a lab wipe under the sampling needle tip.By hand, loosen and remove the joint connected to the 3-way block of the sampling needle flow line.Remove the screws on the upper section of the sampling needle.Be careful not to drop the screws or the holding plate inside the machine during this operation.See fig.5-12.Remove the tube from the black clip.The most probable cause of the reported event was due to operator error.
|