| Model Number |
CATD |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the catd was ovalized approximately 47.0 cm from the hub to the distal tip of the device.Conclusions: evaluation of the returned catd revealed a device with an ovalized distal tip.If the provided peel-away sheath is not used when introducing the catd into a parent device, damage such as an ovalized distal tip may occur.If damage such as this occurs, it may contribute to the reported experienced resistance.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure using an indigo system aspiration catheter d (catd).During the procedure, while inserting a catd into an 8f non-penumbra sheath, the physician experienced resistance; therefore, it was removed.Upon removal, it was noticed that the tip of the catd was pinched.Therefore, the physician removed the sheath and the procedure was completed using a new catd and a new sheath.There was no report of an adverse effect to the patient.
|
| |
|