• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US AVIATOR QUICK TURN SCREWDRIVER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US AVIATOR QUICK TURN SCREWDRIVER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 48770820
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/26/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the distal tip of the quick turn driver broke off intra-op.The surgery was successfully completed using another driver.No adverse consequence to patient was reported.
 
Event or Problem Description
It was reported that the distal tip of the quick turn driver broke off intra-op.The surgery was successfully completed using another driver.No adverse consequence to patient was reported.
 
Additional Manufacturer Narrative
Correction: lot # corrected to 14g999.Method: visual inspection, product history review, complaint history review, nc/capa history review, labelling review, risk assessment result: the reported event was confirmed upon visual inspection of returned product.Manufacturing files were reviewed and no anomalies were found.The instrument was revealed to be approximately 2.5 years old.A materials analysis report on a previous complaint revealed that the hardness and material chemistry were in accordance with the print.Conclusion: the probable root cause of the tip deformation is the age and usage of the instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVIATOR QUICK TURN SCREWDRIVER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key8083001
Report Number0009617544-2018-00271
Device Sequence Number1501896
Product Code LXH
UDI-Device Identifier07613327273045
UDI-Public07613327273045
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number48770820
Device Lot Number14G999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2018
Initial Date Received by Manufacturer 10/26/2018
Supplement Date Received by Manufacturer12/22/2018
Initial Report FDA Received Date11/18/2018
Supplement Report FDA Received Date01/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-