| Catalog Number |
48770820 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the distal tip of the quick turn driver broke off intra-op.The surgery was successfully completed using another driver.No adverse consequence to patient was reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the distal tip of the quick turn driver broke off intra-op.The surgery was successfully completed using another driver.No adverse consequence to patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: lot # corrected to 14g999.Method: visual inspection, product history review, complaint history review, nc/capa history review, labelling review, risk assessment result: the reported event was confirmed upon visual inspection of returned product.Manufacturing files were reviewed and no anomalies were found.The instrument was revealed to be approximately 2.5 years old.A materials analysis report on a previous complaint revealed that the hardness and material chemistry were in accordance with the print.Conclusion: the probable root cause of the tip deformation is the age and usage of the instrument.
|
| |
|