| Model Number |
BIO-COMP SWVLK 3.5X 15.8MM |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was requested/is expected but has not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a mpfl - patella stabilisation.The screw broke at the first turn.The remaining part still lies in the soft tissue and could not be recovered.A swivelock 3.5 was used again.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Updated information obtained (b)(6) 2018: the fragment was not retrieved, it is located in the soft tissue area of the quadriceps tendon and can not move.No x-ray is available.According to the doctor no impairment or consequential damages is foreseeable.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Visual evaluation revealed that the bio-composite screw and the peek eyelet were not returned.Only swivelock driver was returned.Complainant's event is most likely caused by improper bone preparation.
|
| |
|