OGDEN MANUFACTURING PLANT OPTIFLUX F18NRE DIALYZER FINISHED ASSY; DIALYZER, CAPILLARY, HOLLOW FIBER Back to Search Results
Catalog Number 0500308E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
A user facility reported a dialyzer blood leak from a fresenius optiflux f18nre dialyzer that occurred during patient treatment.The blood leak was observed 5 minutes after the start of treatment.The dialyzer and bloodlines were replaced, and the patient completed treatment on the same fresenius 4008s machine with no injury, adverse event, or medical intervention.The dialyzer was discarded and was not available for return.
 
Additional Manufacturer Narrative
The reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was one approved temporary deviation notice (dn) reported on the lot which was unrelated to the complaint event.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX F18NRE DIALYZER FINISHED ASSY
Common Device Name
DIALYZER, CAPILLARY, HOLLOW FIBER
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
475 west 13th street
ogden UT 84404
MDR Report Key8083904
Report Number1713747-2018-00458
Device Sequence Number1771149
Product Code FJI
UDI-Device Identifier00840861100088
UDI-Public00840861100088
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K043244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number0500308E
Device Lot Number18BU01002
Was Device Available for Evaluation? No
Device Age MO
Initial Date Received by Manufacturer 11/05/2018
Supplement Date Received by Manufacturer12/07/2018
Initial Report FDA Received Date11/19/2018
Supplement Report FDA Received Date12/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS 4008S MACHINE; FRESENIUS BLOODLINES; FRESENIUS 4008S MACHINE; FRESENIUS BLOODLINES
Patient Weight67
-
-