Model Number AT120144
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Expiry date (b)(6) 2020.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during preparation for procedure an alleged hair was identified in the sterile packaging.There was no patient contact.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation sealed in the inner sterile pouch.Two photos were also provided by the user for review.A visual inspection of both the returned device packaging and provided photos found a hair sealed within the packaging.Therefore, the investigation is confirmed for the reported issue.The definitive root cause for the hair sealed in the packaging could not be determined based on the available information.However, the hair would have been introduced to the packaging at the manufacturing facility during the packaging process.Operator awareness training was completed, covering personal hairnet adjustment and use, assembly table cleaning, and use of the deionizing gun on all products and pack tubes.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date 02/2020).
 
Event or Problem Description
It was reported that during preparation for procedure an alleged hair was identified in the sterile packaging.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8084095
Report Number2020394-2018-02039
Device Sequence Number1345993
Product Code LIT
UDI-Device Identifier00801741062636
UDI-Public(01)00801741062636
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K120971
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAT120144
Device Catalogue NumberAT120144
Device Lot NumberGFCQ0831
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2018
Initial Date Received by Manufacturer 10/22/2018
Supplement Date Received by Manufacturer03/11/2019
Initial Report FDA Received Date11/19/2018
Supplement Report FDA Received Date03/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-