The reported event that a wrist fusion plate standard bend variax2 120mm was alleged of 'poor fixation' could be confirmed.Based on investigation, the root cause was attributed to be user related.The failure was caused by using the wrong type of screws for the plate selected.Indeed, after inspection, it was found that t10 screws were used distally instead of t8 screws, which explains why they could not be locked into the plate.The head of the t10 screws is bigger than the one of t8.The device inspection revealed the following: it can be noticed based on the deformation of the material of the smartlocking mechanism of the plate that a screw was installed in every circular hole of the plate.This is further confirmed by the x-rays provided.However, none of the oblong holes was filled by a screw.The screws that were returned were tested with the plate.The proximal screws are 3.5, t10 screws which lock perfectly with the plate.The distal screws however, do not lock into the plate at all.These particular screws are variax 1, 2.7 mm screws but the heads are valid for a t10 screwdriver, and this explains the reason they could not be locked.This mistake could have been avoided more easily if variax 2 screws were used, since they are color coded, and thus more easily identified.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
|