Model Number TRUEMETRIX
Medical Device Problem Codes High Test Results (2457); Low Test Results (2458)
Health Effect - Clinical Code Fatigue (1849)
Date of Event 10/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) internal report # (b)(4).Product not returned for evaluation.Customer declined replacement product at this time.Most likely underlying root cause: mlc-20-user's test strip had poor storage(car).Test strip (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure the customer's condition since the initial call and that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for low and high blood glucose results.Father is calling on behalf of the customer.The customer is concerned with results obtained.The expected fasting blood glucose test result range is undisclosed.The customer did report symptom of fatigue.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is not stored by customer according to specification in the car.The test strip lot manufacturer's expiration date is 09/30/2019 and open vial date is undisclosed.The meter memory was not reviewed for previous test result history customer called in states the meter is reading low blood and high blood results.Customer was not feeling well hyperglycemic symptoms of fatigue.Customer declined medical attention.Customer became vulgar and while conducting back to back blood test customer received e-3 twice.Customer states he will discontinue use of meter and supplies and refuse to answer further questions and disconnected call.I will follow up with the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8084252
Report Number1000113657-2018-01181
Device Sequence Number2364517
Product Code NBW
UDI-Device Identifier00021292006051
UDI-Public(01)00021292006051
Combination Product (Y/N)N
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2846
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2018
Initial Report FDA Received Date11/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-