• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH SR HUMERAL HEAD TRIAL 52 X17; SHOULDER JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH SR HUMERAL HEAD TRIAL 52 X17; SHOULDER JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS. Back to Search Results
Catalog Number 5901-S-5217
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
Broken numeral head trial observed intraoperatively.
 
Event or Problem Description
Broken numeral head trial observed intraoperatively.
 
Additional Manufacturer Narrative
The reported event that sr humeral head trial 52 x17 was alleged of 'implant - breakage / deformation - other' could be confirmed.Based on investigation, the root cause was attributed to be user related.The failure was caused by possible inadequate applied force during positioning / removal.Nevertheless more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.Unfortunately no further details were made available.The device inspection (pictures only) revealed the following: the clips for the humeral head trial adaptor are completely broken off.The damages indicate that the surgeon may have encountered some difficulties, if during positioning or removal (using the forked removal tool) is not known though.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.Device was not received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SR HUMERAL HEAD TRIAL 52 X17
Common Device Name
SHOULDER JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS.
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key8084270
Report Number0008031020-2018-00993
Device Sequence Number2792360
Product Code KWS
UDI-Device Identifier07613327100075
UDI-Public07613327100075
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K130895
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5901-S-5217
Device Lot NumberMAC7K22
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/30/2018
Supplement Date Received by Manufacturer11/28/2018
Initial Report FDA Received Date11/19/2018
Supplement Report FDA Received Date12/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-