• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS® SYRINGE MODULE SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS® SYRINGE MODULE SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 10798696
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
 
Event or Problem Description
The customer reported that the filter cracked and leaked during patient use.There was no patient harm.
 
Additional Manufacturer Narrative
The customer¿s report that the filter was cracked was not confirmed; however, the report of a leak was confirmed.Functional testing showed the set leaking between the tubing and the filter outlet and inlet port.The tubing was measured and found to be within specification.The root cause of the leaking from the filter was insufficient solvent application at the engagement between the filter outlet/inlet port and the tubing.
 
Event or Problem Description
The customer reported that the filter cracked and leaked during patient use.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS® SYRINGE MODULE SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key8084317
Report Number9616066-2018-02288
Device Sequence Number1732489
Product Code FPA
UDI-Device Identifier10885403234354
UDI-Public10885403234354
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K811885
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10798696
Device Catalogue Number10798696
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/19/2018
Supplement Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
8015,8110,BD SYRINGE 60ML,TD (B)(6) 2018
-
-