| Model Number |
10798696 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the filter cracked and leaked during patient use.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report that the filter was cracked was not confirmed; however, the report of a leak was confirmed.Functional testing showed the set leaking between the tubing and the filter outlet and inlet port.The tubing was measured and found to be within specification.The root cause of the leaking from the filter was insufficient solvent application at the engagement between the filter outlet/inlet port and the tubing.
|
| |
|
Event or Problem Description
|
|
The customer reported that the filter cracked and leaked during patient use.There was no patient harm.
|
| |
|