| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.(b)(4).Manufacturer contacted customer on (b)(4) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.Daughter is calling on behalf of the customer.The customer is concerned with tests results from results obtained of 44 and 55 mg/dl.Customer's expected fasting blood glucose test result range is 80 - 89 mg/dl.Customer does use alcohol at times on her fingers prior to testing.Customer's fasting result at doctor's office a while ago was 89 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 08/13/2019 and open vial date is two weeks.The meter memory was reviewed for previous test result history (result of 104 mg/dl is daughter's test result and it was not known who the test result of 89 mg/dl belonged to): (b)(6).
|
| |
|
Search Alerts/Recalls
|