| Model Number |
E1450-6 |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the procedure, when the electrode was activated, the tip produces a flame.Generator was used during the event.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the procedure, when the electrode was activated, the tip produced a match size flame.This flame only occurred when the device was activated, so it was self extinguished.Generator was used during the event.Replaced with similar device and it worked fine which completed the procedure.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.The returned electrode did not meet specification as received.Visual inspection found charring on the electrode tip.The reported condition could not be confirmed, but the electrode showed damage consistent with heat damage.Inspection noted that the electrode tip was charred and the coating was beginning to wear off.The device was correctly assembled.Activating the device can result in arcing if other conductive objects are nearby, particularly at a higher power setting.Combustion can occur under these circumstances particularly when an enriched oxygen or nitrous oxide environment is present.The investigation identified the probable root cause of the reported event to be the user activating the device in such a way that the device arced.The instructions for use (ifu) states, both oxygen and nitrous oxide support combustion.Watch for enriched oxygen and nitrous oxide atmospheres near the surgical site, especially during head and neck surgery.Enriched oxygen atmospheres may result in fires and burns to patients or surgical personnel.Do not exceed the maximum power limits as stated in the ifu.Exceeding these power limits may result in patient injury or product damage.If the electrode is damaged, discard it.Always use the lowest power setting to achieve the desired surgical effect.The maximum power settings are detailed in the instructions for use.Some surgeons may elect to buzz the hemostats during a surgical procedure.This practice is not recommended and the hazards of such a practice probably cannot be eliminated.To minimize the risk, when using coated electrodes, place the edge of the electrode against the hemostat or metal instrument to assure a good electrical pathway.Do not activate the generator until the accessory makes contact with the instrument.Additional measures to minimize the risk of performing this technique are listed in the electrosurgical generator users manuals.If information is provided in the future, a supplemental report will be issued.
|
| |
|