• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC XR MP T1 PPS 9X102.5MM; PROSTHESIS HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC XR MP T1 PPS 9X102.5MM; PROSTHESIS HIP Back to Search Results
Model Number N/A
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/25/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that there was lot of cushion debris in the sterile package.Attempts were made to obtain additional information; however, none was available.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Udi : (b)(4).The returned products were visually inspected and the reported event (foam debris in sterile packaging) was confirmed.Device history record (dhr) was reviewed and no discrepancies were found.The likely condition of the product when it left zimmer biomet was conforming to specifications.The root cause of the reported event is likely to be due to transit damage causing the foam to shed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR MP T1 PPS 9X102.5MM
Type of Device
PROSTHESIS HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8085880
MDR Text Key127622067
Report Number0001825034-2018-10690
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 06/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-145090
Device Lot Number6368388
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/25/2018
Initial Date FDA Received11/19/2018
Supplement Dates Manufacturer Received06/05/2019
Supplement Dates FDA Received06/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-