• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD¿ LUER-LOK¿ BLUNT FILL NEEDLE; HYPODERMIC SYRINGE WITH NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD¿ LUER-LOK¿ BLUNT FILL NEEDLE; HYPODERMIC SYRINGE WITH NEEDLE Back to Search Results
Catalog Number 305064
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that before use of the bd¿ luer-lok¿ blunt fill needle that there was moisture on the stopper.There was no report of exposure, serious injury or medical intervention.
 
Additional Manufacturer Narrative
Since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident, therefore a root cause could not be determined.A device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Based on the complaint description it is likely the user is describing visible silicone presence on the stopper.Silicone is an inert, non-toxic medical substance used as a lubricant for disposable hypodermic products.It is an integral part of the syringe, enabling it to perform as required in various clinical applications and does not present a safety or efficacy issue nor does it impact product function.The silicone application process is designed to provide an even distribution of silicone on the interior of the syringe barrel.Silicone has been in use in this application for over 20 years.No reports are known of adverse clinical effects associated with these products and unintentional delivery of silicone fluid lubricant.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that before use of the bd luer-lok blunt fill needle that there was moisture on the stopper.There was no report of exposure, serious injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ LUER-LOK¿ BLUNT FILL NEEDLE
Common Device Name
HYPODERMIC SYRINGE WITH NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
MDR Report Key8085938
Report Number1213809-2018-00788
Device Sequence Number1706044
Product Code FMF
UDI-Device Identifier30382903050643
UDI-Public30382903050643
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K980987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2023
Device Catalogue Number305064
Device Lot Number8218675
Initial Date Received by Manufacturer 10/29/2018
Supplement Date Received by Manufacturer10/29/2018
Initial Report FDA Received Date11/19/2018
Supplement Report FDA Received Date12/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-