Model Number 8225101
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis found no damage or anomalies in the construction of the device that would have resulted in the reported event.The tip fit securely in the handle.The resistance of the assembly was within specification.There was no intermittent behavior while manipulating the tip, connector, and wire.The probe was plugged into a nim system (with a patient simulator) and found that there was no malfunction found as it relates to the complaint.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care professional (hcp) reported via distributer that the probe was not transmitting stimulation during the procedure.Another probe was used and worked normally.There was no patient impact.
 
Additional Manufacturer Narrative
Additional product returned which may have actually been the complaint sample.Fdr is being reported again due to system error.Visually, there was no damage or anomalies in the construction of the device that would have resulted in the reported event.The tip fit securely in the handle.The resistance of the assembly shall measure less than 0.3 ohms while using a 4-wire ohm meter (a 2-wire meter was used in lieu of the 4 wire); the actual measurement was 0.2 ohm which is in specification.There was no intermittent behavior while manipulating the tip, connector, and wire.The probe was plugged into a nim system (with a patient simulator) using 1.0 ma stimulating current and 100uv threshold; when stimulating, the system consistently returned 1.02 ma and a waveform / event tone over 300uv.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM PROBE
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key8086332
Report Number1045254-2018-00609
Device Sequence Number2324895
Product Code ETN
UDI-Device Identifier20681490176686
UDI-Public20681490176686
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K934426
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/06/2024
Device Model Number8225101
Device Catalogue Number8225101
Device Lot Number0212274773
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2018
Supplement Date Received by Manufacturer07/20/2020
Initial Report FDA Received Date11/19/2018
Supplement Report FDA Received Date08/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-