STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS Back to Search Results
Catalog Number 5530-G-311
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Code Injury (2348)
Date of Event 10/23/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding joint instability and poly damage/wear involving a triathlon insert was reported.The instability event was not confirmed, however the damage to the insert was confirmed based on the photograph of the device provided.Method & results: device evaluation and results: no product was returned for evaluation however one photograph was provided.The photograph shows a recently explanted insert with blood on the surface.There is damage/wear on the posterior aspect of the bearing surface on one compartment only.The locking wire is detached from the insert.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusions: no product was returned for evaluation however one photograph was provided.The photograph shows a recently explanted insert with blood on the surface.There is damage/wear on the posterior aspect of the bearing surface on one compartment only.The locking wire is detached from the insert.The exact cause of the event could not be determined because insufficient information was provided.Additional information, including device evaluation, operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
It was reported that patient's right knee was revised due to instability.Intra-operatively, the poly was found to be noticeably damaged on the right side most posterior aspect of the device.Patient's 11 mm cr insert was revised to a 16 mm insert.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Common Device Name
KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8086733
Report Number0002249697-2018-03758
Device Sequence Number16204093
Product Code MBH
UDI-Device Identifier07613327050448
UDI-Public07613327050448
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/30/2017
Device Catalogue Number5530-G-311
Device Lot NumberMLDDT1
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2018
Initial Report FDA Received Date11/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-